Patient Monitor
ZD120D Multi-Parameter Patient Monitor
12.1” high resolution color TFT LCD display. Real time S_T segment analysis, pace-maker detection. Large-font display and ON-line help function Indication of Incorrect manipulation and malfunction analysis. Patient info input management function. Multi-lead ECG wave forms display in phase.Unique accessory management cabinet
Features & Benefits:
6 Standards parameters:ECG,RESP,NIBP,SPO2,1-TEMP, PR/HR
12.1 “ high resolution color TFT LCD display
Real time S_T segment analysis, pace-maker detection.
Large-font display and on-line help function
Multi-lead ECG waveforms display in phase
Large volume of tabular and graphic trends information storage and easy to recall.
Multi-displaying of FHR: number, battery status indicator, Voice intensity indicator.
Capture dynamic waveforms.
Efficient resistance of defibrillator and HF knife.
Up to 2 hours woking capacity of built-in rechargeable battery.
Unique accessory management cabinet.
Networking capacity.
Options of IBP and EtCO2.
Safety:
IEC60601-1 approved, Category IBF, CF Type
Dimension and Weight:
Dimension: 300mm X 400mm X 500mm
G. weight:<9.0 KGS
Operation Environment
Dimension: 300mm X 400mm X 500mm
Temperature: working 0-40'C
Transporation and storage -20+- 60'C
Humidity: working <85%
Power: AC 100-240V, 50/60 Hz
Patient range: Neonate,Pediatric, and adult patient
Performance Specifications:
Display:
12.1" color TFT. Rolling and refreshing waveforms display.
Resolution 800X600
Large font display
Satandard display.
Multi channels display
Patient range: Neonate,Pediatric, and adult patient
Trace: 6 waveforms
Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
Indicator: Power battery indicator light, QRS beep and alarm sound
Battery: Rechargeable lead acid cell, 12V/4AH, MAx 24 hours for charging, 2 hours for continuous working.
Tread time: above 72 hours

Alarm: User adjustable High, Medium Nd Low limits, 3-levelAudible and visual alarm
Networking: Connected to central monitoring system
Recorder: Built-in,thermal array,plethysmogram waveforms: 2 Channels, Record mode:manual, on alarm, time defined.
Recording width:50mm
Printing Speed:25mm/s
Recording type: Frozen waveforms record, NIBP recall record, trend table record, Alarm record, fixed-time record
Specifection
ECG:
Lead mode: 5 - lead (R,L,F,N,C)
Lead selection: I, II, III,avR, avL,avF,V
Waveform:2 and 6 channel selectable
Gain selection: 0.25mm/mv, 0.5mm/mv, 1mm/mv,2mm/mv
Sweep Spead:12.5mm/s; 15mm/s; 50mm/s
Hearl Rate Range: Adult:15-300bpm; Neonate:/Pediatric:15-350bpm.
Accuracy:+1bpm or +1 %, whichever is greater.
Resolution: 1bpm
Filter:Surgery mode:1-20 Hz, Monitor mode:0.5-40Hz, Daignostic Mode:0.05-130Hz
Scalling Signal:1mV,+_3%
Protection:Withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
Alarm range:15-350bpm
S-T segnent detection:Measurement range:-2.0mV- +2.0 mV, Alarm Range:-2.0mV- +2.0 mV, Accuracy:-0.8mV-+0.8mV, Error:+_0.02mV
Arrhythmia analysis:Yes

NIBP:
Method : Digital Automatic oscillometric
Operation mode: Manual/Automatic/Continuous
Auto measurement time: Adjustable(1-480min)
Measurement Unit: mmHg/Kpa selectable
Measurement Types: Systolic, Diastolic, Mean
Measurement Range:
Range of Systolicc pressure: Adult:40-270mmHg, Pediatric: 40-200mmHg, Neonate:40-135mmHg.
Range of Mean pressure: Adult:20-235mmHg, Pediatric: 20-165mmHg, Neonate:20-110mmHg.
Range of diastolic pressure: Adult:10-215mmHg, Pediatric: 10-150mmHg, Neonate:10-100mmHg.
Over-pressure protection: Double safety protection
Resolution: 1mmHg
Alarm: Systolicc, diastolic, Mean
SPO2:
Measurement range: 0-100%
Resolution :1%.
Accuracy : 2%(70-100%); 0-69% unspecified
Alarm range : 0-100%
Pulse Rate: Range:20-300bpm
Resolution:1 Bpm
Error:1bpm or 2%, whichever is greater
RESPIRATION :
Method: Thoracic impedance
Measurement Range: Adult: 7- 120 rpm; Neonate/Pediatric:7-150rpm
Apnea alarm: YES, 10-40s
Resolution: 1 rpm
Accuracy : 2rpm
TEMPERATURE :
Compatible Probe: YSI or CYF
Measurement Range: 5-50 c
Resolution:  0.1 c
Accuracy: 0.1 c
Channel: dual-channel, provide T1, T2, DT
Refreshing Time: about 1s
Average measuring time: <10s
IBP :
Measurement range: -10-300mmHg
Pressure name: ART, PA, CVP, RAP, LAP,ICP,P2
Channel: 2 channel
Pressure  transducer:Sensitivity: 5 u V/V/mmHg
Impedance range:300-3000
Accuracy: 1mmHg or 2%, whichever is greater(exclusive of transducer)
Alarm range: -10-300mmHgff
EtCO2 :
Sidestream CO2
Measurement range: 0-99mmHg
Accuracy: 2mmHg(0-40mmHg)
Sample Range:100ml/min
Sampling rate accuracy:15%
Respiration rate:0-120rpm
Respiration accuracy: 2rpm(0-70rpm), 5rpm(>70rpm)
Respiration time: <240msec (10% to 90%)
Delay Time: <2sStandard Configuration:ECG, NIBP,RESP,1-TEMP,SPO2,PR
Optional: Thermal Recorder, 1-IBP, 2-IBP,ETCO2